Geometrically-derived density compensation function for gridding reconstruction of 2D and 3D non-Cartesian MRI data
{{output}}
Purpose: Density compensation function (DCF) is essential for minimizing reconstruction errors in non-Cartesian MRI. This study aimed to extend a geometrically-derived DCF (gDCF), previously validated for 2D radial MRI, to arbitr... ...