首页 正文

Scientific reports. 2025 May 30;15(1):18971. doi: 10.1038/s41598-025-03743-9 Q13.82024

Accuracy of the HeartMate 3 Left Ventricular Assist Device Flow Estimation

HeartMate 3左心室辅助装置流量估算的准确性 翻译改进

Theodor Abart  1, Marko Grujic  1, Philipp Aigner  1, Michael Röhrich  2, Stefan Jakubek  3, Daniel Zimpfer  1, Marcus Granegger  4

作者单位 +展开

作者单位

  • 1 Christian Doppler Laboratory for Mechanical Circulatory Support, Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria.
  • 2 Department of Anesthesia, Intensive Care Medicine and Pain Medicine, Medical University of Vienna, Vienna, Austria.
  • 3 Division of Control and Process Automation, Institute of Mechanics and Mechatronics, Vienna University of Technology, Vienna, Austria.
  • 4 Christian Doppler Laboratory for Mechanical Circulatory Support, Department of Cardiac and Thoracic Aortic Surgery, Medical University of Vienna, Vienna, Austria. marcus.granegger@meduniwien.ac.at.
  • DOI: 10.1038/s41598-025-03743-9 PMID: 40447681

    摘要 中英对照阅读

    Left ventricular assist devices (LVADs), such as the HeartMate 3 (HM3), are a treatment option for advanced heart failure (HF). Non-invasive monitoring of heart-pump interaction is crucial but relies on estimated parameters only. Despite the widespread clinical utilization of HM3 estimated flow rate (Qest) and pulsatility index (PI) their accuracy was not yet quantified. This study assessed the HM3 monitoring by describing underlying physical characteristics, estimation algorithms and deriving accuracy of estimates in clinical practice. Identification measurements were performed to quantify the physical characteristics and mechanisms of flow estimation. Virtual patients coupled to a hybrid hock circulatory loop were employed to evaluate estimator accuracies in realistic operating conditions. The correlation between Qest and torque-generating-current was strong (r² > 0.99, p < 0.001), confirming that estimation is based on speed, current, and viscosity. Also, the PI is directly derived from current. However, the underlying current/flow relationship is ambiguous (one current value corresponds to two flow rates). The HM3 flow estimator showed an RMSE of 1.63 L/min and a correlation coefficient of r = 0.86 (p < 0.001). The non-linear relationship obscures the systolic portion of flow as current declines during peak systole at higher flow rates. HM3 monitoring is unable to accurately represent the actual hemodynamics, particularly in the transition between full support and partial support, which is a region highly prevalent and crucial in clinical practice. It is therefore recommended to interpret values and trends of the HM3 monitoring system with caution.

    Keywords:left ventricular assist device; flow estimation

    左心室辅助装置(如HeartMate 3,HM3)是治疗晚期心脏衰竭的一种选择。对心脏泵相互作用的非侵入性监测至关重要,但目前仅依赖于估算参数。尽管临床广泛使用了HM3估计流量率(Qest)和脉动指数(PI),但是其准确性尚未量化。本研究通过描述物理特性、估计算法并推导出实际操作中的估计准确度来评估HM3监测系统。进行识别测量以量化物理特性和流体估算机制。利用与混合霍克循环耦合的虚拟患者,在真实的运行条件下评估估计器精度。Qest和产生扭矩电流之间的相关性很强(r² > 0.99,p < 0.001),这证实了估计是基于速度、电流和粘度进行的。此外,PI直接由电流得出。然而,基础的电流/流量关系模糊不清(一个电流值对应两个流量率)。HM3流体估算器显示均方根误差为1.63 L/min,并且相关系数r = 0.86(p < 0.001)。非线性关系在高流量率情况下,心缩期峰值时电流下降期间遮蔽了收缩期的流量。HM3监测系统无法准确反映实际血流动力学情况,尤其是在全支持和部分支持之间的过渡区,这是一个临床实践中普遍存在且至关重要的区域。因此建议谨慎解读HM3监测系统的值和趋势。

    © 2025. The Author(s).

    关键词:左心室辅助设备; 流量估计

    翻译效果不满意? 用Ai改进或 寻求AI助手帮助 ,对摘要进行重点提炼
    Copyright © Scientific reports. 中文内容为AI机器翻译,仅供参考!

    相关内容

    期刊名:Scientific reports

    缩写:SCI REP-UK

    ISSN:2045-2322

    e-ISSN:

    IF/分区:3.8/Q1

    文章目录 更多期刊信息

    全文链接
    引文链接
    复制
    已复制!
    推荐内容
    Accuracy of the HeartMate 3 Left Ventricular Assist Device Flow Estimation