International journal of computer assisted radiology and surgery. 2022 Jan;17(1):15-26. doi: 10.1007/s11548-021-02484-0 Q32.32024

An image-guided hybrid robot system for dental implant surgery

一种用于牙科植入手术的图像引导型混合机器人系统 翻译改进

Yuan Feng  1, JiaCheng Fan  1, BaoXin Tao  2, ShiGang Wang  1, JinQiu Mo  1, YiQun Wu  2, QingHua Liang  3, XiaoJun Chen  4

作者单位 +展开

作者单位

  • 1 School of Mechanical Engineering, Shanghai Jiao Tong University, Room 901, Dongchuan Road 800, Minhang District, Shanghai, 200240, China.
  • 2 Department of Second Dental Center, Shanghai Ninth People's Hospital, College of Stomatology, Shanghai Jiao Tong University, School of Medicine, National Clinical Research Center for Oral Disease, Shanghai Key Laboratory of Stomatology and Shanghai Research Institute of Stomatology, 639, Zhizaoju Road, Shanghai, 200011, China.
  • 3 School of Mechanical Engineering, Shanghai Jiao Tong University, Room 901, Dongchuan Road 800, Minhang District, Shanghai, 200240, China. qhliang@sjtu.edu.cn.
  • 4 School of Mechanical Engineering, Shanghai Jiao Tong University, Room 901, Dongchuan Road 800, Minhang District, Shanghai, 200240, China. xiaojunchen@sjtu.edu.cn.
  • DOI: 10.1007/s11548-021-02484-0 PMID: 34449036

    摘要 Ai翻译

    Purpose: Dental implant surgery is an effective method for remediating the loss of teeth. Robot is expected to increase the accuracy of dental implant surgery. However, most of them are industrial serial robot, with low stiffness and non-unique inverse kinematic solution, which may reduce the success rate and safety of robotic surgery. Compared to serial robot, parallel robot is more stiffness and has unique inverse kinematic. However, its workspace is small, which may not meet surgical requirements. Therefore, a novel hybrid robot dedicated to dental implant is proposed.

    Methods: The hybrid robot is composed of three translation joints, two revolute joints, and Stewart parallel manipulator. Stewart is used for performing surgical operation, while the joints are used for enlarging the workspace of Stewart. In order to ensure the safety of robot motion, physical human-robot interaction based on a variable admittance controller is applied in the robotic system. In addition, considering the small workspace of Stewart, an optimal model is proposed to minimize the joint movement of Stewart in adjusting the orientation of drill bit.

    Results: Phantom experiments were carried out based on the prototype robot. In the experiments, the optimal model could be solved after 20 iterations, finding an ideal joint configuration. The proposed variable admittance controller could enhance comfort level effectively. The accuracy of robot is evaluated by angle, entry and exit deviation, which are 0.74 ± 0.25°, 0.93 ± 0.28 mm, and 0.96 ± 0.23 mm, respectively.

    Conclusion: The phantom experiments validate the functionality of the proposed hybrid robot. The satisfactory performance makes it more widely used in the practical dental implant surgery in the future.

    Keywords: Dental implant surgery; Hybrid robot system; Physical human–robot interaction; Stewart manipulator.

    Keywords:image-guided system; dental implant surgery

    Copyright © International journal of computer assisted radiology and surgery. 中文内容为AI机器翻译,仅供参考!

    相关内容

    期刊名:International journal of computer assisted radiology and surgery

    缩写:INT J COMPUT ASS RAD

    ISSN:1861-6410

    e-ISSN:1861-6429

    IF/分区:2.3/Q3

    文章目录 更多期刊信息

    全文链接
    引文链接
    复制
    已复制!
    推荐内容
    An image-guided hybrid robot system for dental implant surgery