A case of a C-stem fracture at the head-neck junction and a review of the literature
{{output}}
We report the first case of a fracture of the standard C-stem in combination with a large metal-on-metal articulation. This occurred at the head-neck junction. Analysis of the fractured stem showed evidence of fatigue failure with possible corrosion. The use o... ...